Grant: Research Tools and Instruments - Category 1 (<$150,000)
McMaster University
$121,960.00
Upgrade of a fluorescence lifetime measurement system
FY 2010-11
Summary
This proposal requests an upgrade of an existing time-domain fluorescence system through the acquisition of a synchroscan sweeping module for a steak camera in the system and a new picoseconds diode laser at 440 nm. The requested upgrade is essential to support research projects in developing technologies for high content screening of cancer drug leads and minimally-invasive optical biopsy guided cancer intervention.
The current configuration (laser: 0-1 KHz, 355 nm, 300 ps; detection: 0-4 MHz) of the streak camera based time-domain fluorescence system is designed for single point, time-resolved spectroscopic applications and proof-of-principle study of fluorescence lifetime imaging applications. To further extend our research into live cell imaging and in-vivo diagnosis applications, significant improvement of the repetition rate and data acquisition speed is required. For example, a full 2D fluorescence lifetime image needs 900 seconds data acquisition time with the current configuration, which is impractical for live cell imaging. The most significant limitation of the current configuration is the low repetition rate (1 KHz laser & 4 MHz detection) and excitation at 355 nm. The proposed upgrade with a synchroscan module and a blue picosecond diode laser allows up to 50 times improvement in data acquisition for live cell imaging applications. This upgrade will also improve the temporal resolution of the streak camera system from 50 ps to 2 ps, which allows it to be used as a high speed photo detection system in a number of research projects at McMaster such as characterization of microfabricated high speed photo detector arrays and picoseconds pulses in laser micromachining.
The team of researchers has extensive experiences in high speed photo detection systems including streak camera. The modular design of the system allows easy access and sharing between research projects. Graduate students and postdoctoral fellows will not only benefit from the uninterrupted progress in their research projects but also from hands-on experiences on a sophisticated, high sensitivity, high speed photo detection system as well as the expertise in high speed time-domain measurements.
Program
Research Tools and Instruments - Category 1 (<$150,000)
Recipient
Location
Ontario, Canada
Recipient Type
Academia
Type
Grant
Source
NSERC's Awards Data
As published
Application id
402121-2011
Name nom
Fang, Qiyin
Department département
Engineering Physics
Organization id
27
Institution établissement
McMaster University
Province en
Ontario
Province fr
Ontario
Country en
CANADA
Country fr
CANADA
Fiscal year exercice financier
2010
Competition year année de concours
2011
Award amount
121960
Program id
EQPEQ
Program name en
Research Tools and Instruments - Category 1 (<$150,000)
Program name fr
Outils et instruments de recherche - catégorie 1 (<150 000$)
Group en
Discovery grants programs
Group fr
Programmes de subventions à la découverte
Committee code
1510
Committee name en
Electrical and Computer Engineering
Committee name fr
Génie électrique et informatique
Area of application code
1102
Area of application group en
Health, education and social services
Area of application group fr
Santé publique, éducation et services sociaux
Area of application en
Biomedical engineering
Area of application fr
Génie biomédical
Research subject code
1900
Research subject group en
Biomedical engineering
Research subject group fr
Génie biomédical
Research subject en
Biomedical engineering
Research subject fr
Génie biomédical
Partie
1
Nb partie
1
Application title
Upgrade of a fluorescence lifetime measurement system
Keyword
NSERC
Application summary
This proposal requests an upgrade of an existing time-domain fluorescence system through the acquisition of a synchroscan sweeping module for a steak camera in the system and a new picoseconds diode laser at 440 nm. The requested upgrade is essential to support research projects in developing technologies for high content screening of cancer drug leads and minimally-invasive optical biopsy guided cancer intervention.
The current configuration (laser: 0-1 KHz, 355 nm, 300 ps; detection: 0-4 MHz) of the streak camera based time-domain fluorescence system is designed for single point, time-resolved spectroscopic applications and proof-of-principle study of fluorescence lifetime imaging applications. To further extend our research into live cell imaging and in-vivo diagnosis applications, significant improvement of the repetition rate and data acquisition speed is required. For example, a full 2D fluorescence lifetime image needs 900 seconds data acquisition time with the current configuration, which is impractical for live cell imaging. The most significant limitation of the current configuration is the low repetition rate (1 KHz laser & 4 MHz detection) and excitation at 355 nm. The proposed upgrade with a synchroscan module and a blue picosecond diode laser allows up to 50 times improvement in data acquisition for live cell imaging applications. This upgrade will also improve the temporal resolution of the streak camera system from 50 ps to 2 ps, which allows it to be used as a high speed photo detection system in a number of research projects at McMaster such as characterization of microfabricated high speed photo detector arrays and picoseconds pulses in laser micromachining.
The team of researchers has extensive experiences in high speed photo detection systems including streak camera. The modular design of the system allows easy access and sharing between research projects. Graduate students and postdoctoral fellows will not only benefit from the uninterrupted progress in their research projects but also from hands-on experiences on a sophisticated, high sensitivity, high speed photo detection system as well as the expertise in high speed time-domain measurements.