Grant: Discovery Grants Program - Individual
University of Saskatchewan
$21,000.00
Design and Testing Gradient Free Magnetic Resonance Imaging
EF 2020-21
Résumé
The progress made in developing a new generation of gradient-free magnetic resonance imagers (MRIs) will continue. Progress made includes the design of a wrist-sized MRI for the International Space Station (ISS), for the Canadian Space Agency (CSA), and the construction and testing of two wrist-sized gradient-free MRIs.
At the heart of gradient-free MRI is a novel radio frequency (RF) transmitting coil. The RF field design of the coil allows for direct spatial encoding with no need for magnetic field gradients - gradient-free MRI. When combined with a low field (500 gauss), a very portable MRI results. The magnet mass of both wrist MRIs constructed to date is less than 5 kg. The spatial variation of both RF magnitude and RF phase can be used for image encoding. One of my existing wrist MRIs combines a varying RF field with motorized motion of the RF coils. My other existing MRI uses spatially varying RF phase to achieve spatial encoding. The two MRI designs will be developed further, along with other designs.
The research will be done by HQP trainees following the very successful approach, to date, of doing mathematically based research projects coupled with investigations using the wrist-size MRIs. Modification of the existing MRIs and construction of one more wrist-sized MRI is anticipated. A start-up company has been planned to build helmet-size gradient-free MRIs for use in existing remote-presence enabled clinics in the Canadian north. But our focus for the NSERC sponsored research will be wrist-sized MRIs because our purpose is to develop the MRI engineering and not to study anatomy or physiology. This approach also keeps the hardware costs down.
The ISS-MRI design uses a highly homogeneous magnetic field and a linearly varying transmitted RF phase profile. The proposed work uses non-linearly varying transmitted RF phase profiles. A non-linearly varying phase profile means that the Fourier transform cannot be used directly to reconstruct the image from the received RF signal, except in a very small field of view near the center of the RF transmit coil. Novel spin echo pulse sequences, using composite pulses, are also required. Our wrist-size MRIs also do not use homogeneous magnetic fields so that isomagnetic surfaces in the object are excited to give non-planar slices. These new approaches require mathematical transformations that go beyond the standard Fourier transformation for image reconstruction. Mathematical reconstruction algorithms will be the focus of some of the research projects. Other research projects will involve RF coil design, RF pulse sequence design and magnetic field design. LT Imaging and MRI Tech will provide the MRI hardware needed for the projects.
Portable gradient-free MRI has the potential to displace both x-ray and ultrasound as medical imaging modalities. The proposed research will move MRI technology toward those future roles to deliver accessible medical diagnostics.
Programme
Discovery Grants Program - Individual
Bénéficiaire
Emplacement
Saskatchewan, Canada
Type de bénéficiaire
Academia
Type
Grant
Source
NSERC's Awards Data
Tel que publié
Application id
RGPIN-2017-03740
Name nom
Sarty, Gordon
Department département
Biomedical Engineering/Psychology
Organization id
16
Institution établissement
University of Saskatchewan
Province en
Saskatchewan
Province fr
Saskatchewan
Country en
CANADA
Country fr
CANADA
Fiscal year exercice financier
2020
Competition year année de concours
2017
Award amount
21000
Program id
RGPIN
Program name en
Discovery Grants Program - Individual
Program name fr
Programme de subventions à la découverte - individuelles
Group en
Discovery grants programs
Group fr
Programmes de subventions à la découverte
Committee code
1505
Committee name en
Physics
Committee name fr
Physique
Area of application code
1102
Area of application group en
Health, education and social services
Area of application group fr
Santé publique, éducation et services sociaux
Area of application en
Biomedical engineering
Area of application fr
Génie biomédical
Research subject code
1901
Research subject group en
Biomedical engineering
Research subject group fr
Génie biomédical
Research subject en
Biomedical technology
Research subject fr
Technologie biomédicale
Installment
inst:4-5
Partie
2
Nb partie
5
Application title
Design and Testing Gradient Free Magnetic Resonance Imaging
Keyword
NSERC
Application summary
The progress made in developing a new generation of gradient-free magnetic resonance imagers (MRIs) will continue. Progress made includes the design of a wrist-sized MRI for the International Space Station (ISS), for the Canadian Space Agency (CSA), and the construction and testing of two wrist-sized gradient-free MRIs.
At the heart of gradient-free MRI is a novel radio frequency (RF) transmitting coil. The RF field design of the coil allows for direct spatial encoding with no need for magnetic field gradients - gradient-free MRI. When combined with a low field (500 gauss), a very portable MRI results. The magnet mass of both wrist MRIs constructed to date is less than 5 kg. The spatial variation of both RF magnitude and RF phase can be used for image encoding. One of my existing wrist MRIs combines a varying RF field with motorized motion of the RF coils. My other existing MRI uses spatially varying RF phase to achieve spatial encoding. The two MRI designs will be developed further, along with other designs.
The research will be done by HQP trainees following the very successful approach, to date, of doing mathematically based research projects coupled with investigations using the wrist-size MRIs. Modification of the existing MRIs and construction of one more wrist-sized MRI is anticipated. A start-up company has been planned to build helmet-size gradient-free MRIs for use in existing remote-presence enabled clinics in the Canadian north. But our focus for the NSERC sponsored research will be wrist-sized MRIs because our purpose is to develop the MRI engineering and not to study anatomy or physiology. This approach also keeps the hardware costs down.
The ISS-MRI design uses a highly homogeneous magnetic field and a linearly varying transmitted RF phase profile. The proposed work uses non-linearly varying transmitted RF phase profiles. A non-linearly varying phase profile means that the Fourier transform cannot be used directly to reconstruct the image from the received RF signal, except in a very small field of view near the center of the RF transmit coil. Novel spin echo pulse sequences, using composite pulses, are also required. Our wrist-size MRIs also do not use homogeneous magnetic fields so that isomagnetic surfaces in the object are excited to give non-planar slices. These new approaches require mathematical transformations that go beyond the standard Fourier transformation for image reconstruction. Mathematical reconstruction algorithms will be the focus of some of the research projects. Other research projects will involve RF coil design, RF pulse sequence design and magnetic field design. LT Imaging and MRI Tech will provide the MRI hardware needed for the projects.
Portable gradient-free MRI has the potential to displace both x-ray and ultrasound as medical imaging modalities. The proposed research will move MRI technology toward those future roles to deliver accessible medical diagnostics.