Grant: Engage Grants Program
University of Western Ontario
$24,950.00
Optimization and manufacturability analysis of a catheter delivery system for piezo-resistive pressure sensor
EF 2011-12
Résumé
The industrial partner of the current proposal, Scisense Inc., represents one of the market leaders in the area of
micro-sensing devices for medical research applications. They have successfully developed and marketed a line
of intra cardiac pressure and volume measurement devices for mice. The company aims to broaden its current
line of products by launching enhanced catheter delivery systems specifically intended for medium and large
rodents.
However, Scisense's expansion plans in this direction are currently hindered by product development
challenges associated with the new line of products. The piezo-resistive pressure sensor in the catheter can give
false readings because it is influenced by deflection of its housing as the catheter is routed through the animal.
Additionally, differences in rates of thermal expansion of the components can induce thermal stresses in the
sensor again resulting in false readings. Scisense lacks the expertise to conduct detailed optimization analysis
and with micro manufacturing which will be required to fabricate a more robust, yet compact device.
The project is focused on analyzing and optimizing potential design configurations with the objective of
reducing the forces imparted on the sensor to an acceptable value. The stiffness of the supporting structure
must be maximized however severe size restrictions are imposed since it must be able to pass through the
animal's circulatory system without significant trauma. A select few designs will be chosen for fabrication to
validate the models. At the same time, the collaborator, Dr. Tutunea-Fatan, will assess the manufacturability of
the designs. Validation will be accomplished by initially performing loading experiments on the prototype
housings to verify their deflection characteristics. Then the company will assemble the housings with sensors
and verify the device performance through a series of standard tests.
Programme
Engage Grants Program
Bénéficiaire
Emplacement
Ontario, Canada
Type de bénéficiaire
Academia
Type
Grant
Source
NSERC's Awards Data
Tel que publié
Application id
412459-2011
Name nom
Salisbury, Shaun
Department département
Mechanical and Materials Engineering
Organization id
36
Institution établissement
University of Western Ontario
Province en
Ontario
Province fr
Ontario
Country en
CANADA
Country fr
CANADA
Fiscal year exercice financier
2011
Competition year année de concours
2011
Award amount
24950
Program id
EGP
Program nane en
Engage Grants Program
Program name fr
Programme de subventions d'engagement partenariat
Group en
Research partnerships programs
Group fr
Programmes de partenariat
Committee code
1553
Committee name en
Ontario Internal Decision Committee
Committee name fr
Comité de décision interne pour l'Ontario
Area of application code
518
Area of application group en
Manufacturing processes and products
Area of application group fr
Procédés et produits industriels
Area of application en
Medical equipment and apparatus
Area of application fr
Appareils et instruments médicaux
Research subject code
2107
Research subject group en
Mechanical engineering
Research subject group fr
Génie mécanique
Research subject en
Modelling, simulation and finite element methods
Research subject fr
Modélisation, simulation et méthodes des éléments finis
Partie
1
Nb partie
1
Application title
Optimization and manufacturability analysis of a catheter delivery system for piezo-resistive pressure sensor
Keyword
NSERC
Application summary
The industrial partner of the current proposal, Scisense Inc., represents one of the market leaders in the area of
micro-sensing devices for medical research applications. They have successfully developed and marketed a line
of intra cardiac pressure and volume measurement devices for mice. The company aims to broaden its current
line of products by launching enhanced catheter delivery systems specifically intended for medium and large
rodents.
However, Scisense's expansion plans in this direction are currently hindered by product development
challenges associated with the new line of products. The piezo-resistive pressure sensor in the catheter can give
false readings because it is influenced by deflection of its housing as the catheter is routed through the animal.
Additionally, differences in rates of thermal expansion of the components can induce thermal stresses in the
sensor again resulting in false readings. Scisense lacks the expertise to conduct detailed optimization analysis
and with micro manufacturing which will be required to fabricate a more robust, yet compact device.
The project is focused on analyzing and optimizing potential design configurations with the objective of
reducing the forces imparted on the sensor to an acceptable value. The stiffness of the supporting structure
must be maximized however severe size restrictions are imposed since it must be able to pass through the
animal's circulatory system without significant trauma. A select few designs will be chosen for fabrication to
validate the models. At the same time, the collaborator, Dr. Tutunea-Fatan, will assess the manufacturability of
the designs. Validation will be accomplished by initially performing loading experiments on the prototype
housings to verify their deflection characteristics. Then the company will assemble the housings with sensors
and verify the device performance through a series of standard tests.