Canada Spends Logo

Explorez notre analyse du budget 2026

Grant: Idea to Innovation

Queen's University

$116,000.00

Fiber optic absorption detector for micro-chemical analysis

EF 2009-10

Résumé
We will develop a UV-Vis absorption detector that is capable of quantifying dilute chemical substances at four wavelengths between 480 -255 nm and in volumes of less than 10 nanoliters. The detector will be an alternative to fluorescence detection in micro-analytical separation methods, such as micro-HPLC, microfluidic devices and capillary electrophoresis. This alternative is needed, because compounds that do not exhibit native fluorescence presently require an error-prone and time-consuming labeling step. Furthermore, the new detector is also expected to be a viable alternative to absorption detectors commonly used in commercial chemical analysis labs. At present, many HPLC systems are coupled to absorption detectors that achieve excellent sensitivity and detection limits for detection volumes in the micro-liter range. By reducing the size of the separation system to micro-HPLC (nanoliter detection volumes), a considerable amount of money could be saved, but this is not possible with present absorption detectors. We expect that our detector will reduce the analysis time and the environmental footprint (solvent consumption) of the chemical analysis, while retaining the sensitivity, accuracy and detection limits of analytical scale HPLC. By the conclusion of Phase I we will have developed and built at least one prototype of a 4-wavelength absorption detector. This detector will be tested along-side commercial absorption detectors in an analysis lab in the pharmaceutical industry. One graduate student will be involved in the research and development of the detector and will gain valuable experience in dealing with commercial clients.
Programme
Idea to Innovation
Bénéficiaire
Emplacement
Ontario, Canada
Type de bénéficiaire
Academia
Type
Grant
Source
NSERC's Awards Data

Tel que publié

Application id
387147-2009
Name nom
Loock, HansPeter
Department département
Chemistry
Organization id
29
Institution établissement
Queen's University
Province en
Ontario
Province fr
Ontario
Country en
CANADA
Country fr
CANADA
Fiscal year exercice financier
2009
Competition year année de concours
2009
Award amount
116000
Program id
I2IPJ
Program nane en
Idea to Innovation
Program name fr
De l'idée à l'innovation
Group en
Research partnerships programs
Group fr
Programmes de partenariat
Committee code
96
Committee name en
RPP Internal Decision Cttee
Committee name fr
Cté de décision interne - PPR
Area of application code
515
Area of application group en
Manufacturing processes and products
Area of application group fr
Procédés et produits industriels
Area of application en
Instrumentation technology
Area of application fr
Technologie de l'instrumentation
Research subject code
3505
Research subject group en
Analytical chemistry
Research subject group fr
Chimie analytique
Research subject en
Analytical spectroscopy
Research subject fr
Spectroscopie analytique
Partie
1 & 2
Nb partie
2
Application title
Fiber optic absorption detector for micro-chemical analysis
Keyword
NA
Application summary
We will develop a UV-Vis absorption detector that is capable of quantifying dilute chemical substances at four wavelengths between 480 -255 nm and in volumes of less than 10 nanoliters. The detector will be an alternative to fluorescence detection in micro-analytical separation methods, such as micro-HPLC, microfluidic devices and capillary electrophoresis. This alternative is needed, because compounds that do not exhibit native fluorescence presently require an error-prone and time-consuming labeling step. Furthermore, the new detector is also expected to be a viable alternative to absorption detectors commonly used in commercial chemical analysis labs. At present, many HPLC systems are coupled to absorption detectors that achieve excellent sensitivity and detection limits for detection volumes in the micro-liter range. By reducing the size of the separation system to micro-HPLC (nanoliter detection volumes), a considerable amount of money could be saved, but this is not possible with present absorption detectors. We expect that our detector will reduce the analysis time and the environmental footprint (solvent consumption) of the chemical analysis, while retaining the sensitivity, accuracy and detection limits of analytical scale HPLC. By the conclusion of Phase I we will have developed and built at least one prototype of a 4-wavelength absorption detector. This detector will be tested along-side commercial absorption detectors in an analysis lab in the pharmaceutical industry. One graduate student will be involved in the research and development of the detector and will gain valuable experience in dealing with commercial clients.