Résumé
Engineering advances over the last 30 years have made magnetic resonance imaging (MRI) a powerful imaging modality for the non-invasive investigation of the human body. Routine clinical MRI measures signal from the hydrogen (1H) nuclei in water. However, with the appropriate hardware and imaging techniques, MRI can also acquire signal from other nuclei such as sodium (23Na) and potassium (39K), ions which may be more specific to aspects of tissue injury and disease than the ubiquitous water molecule. The ability to image sodium can provide a measure of cartilage degradation in osteoarthritis, a debilitating condition affecting millions of Canadians. Potassium is a key ion in brain function, but methods to measure it in human brain are lacking. The purpose of this grant is to develop highly sensitive radiofrequency (RF) coil hardware (e.g. sensitive phased arrays) and imaging methodology to enable quantitative imaging of these nuclei either in cartilage (23Na) or brain (39K). This proposal builds on our decade's worth of sodium MRI research, which produced world-leading images of human brain and then knee (funded over last 5 years by NSERC).
Sodium (and potassium) MRI is very challenging because of low concentration in tissue, small magnetogyric ratio, rapid signal decay, complex spin physics, and the need for nucleus-specific hardware and optimal methods. Our previous sodium MRI research on cartilage of the knee focused on the design of optimal acquisition strategies, but we used only standard volume RF coils. However, more complex and sensitive phased-array RF coils, which consist of many small localized elements, can dramatically increase signal-to-noise ratio and thus yield major gains in image resolution and quantification. While phased-array technology is now standard for regular 1H MRI, it is in its infancy for sodium MRI and needs to be explored not only for knee, but also for other body regions such as wrist, ankle, hip, and spine (where cartilage is also affected by osteoarthritis). Potassium has even greater technical challenges to overcome than sodium, and there are only two recent preliminary reports of potassium MRI in human brain.
The novel MRI hardware and methods will be designed, simulated, constructed, programmed, tested, and published by the trainees. The developments will be on a `triple strength' high field 4.7T MRI to enable major gains in signal. We hypothesize that our technical MRI advances will enable the accurate and precise measurement of sodium in cartilage and potassium in brain.
Specific Aims:
1) To develop phased-array receive-only sodium RF coils and dual-frequency (23Na/1H) detunable, concentric transmit volume coils for imaging the knee, ankle, and wrist. In this case the volume coil provides uniform excitation and the phased-array provides high receive sensitivity. The capability for simultaneous reception from both the phased-array and volume coils using counter-rotating-current coil elements will be explored to facilitate accelerated image intensity correction related to the spatially varying RF sensitivity profiles of the phased-array elements.
2) To develop transceive dual-frequency phased-array RF coils for sodium MRI of the spine and hip, where uniform volume excitation is not feasible for sodium. An associated aim is to develop and optimize multiple coil element transmission for uniform excitation.
3) To develop quantitative potassium MRI of human brain.
After 5 years, we will have developed novel multi-nuclear radiofrequency coil hardware and optimized MRI acquisition to enable imaging of (i) sodium in cartilage of the extremities, hip, and spine which are key structures affected in osteoarthritis, and (ii) potassium in the brain, a novel biomarker of brain metabolism.
Programme
Discovery Grants Program - Individual
Bénéficiaire
Emplacement
Alberta, Canada
Type de bénéficiaire
Academia
Type
Grant
Source
NSERC's Awards Data
Tel que publié
Application id
RGPIN-2014-03966
Name nom
Beaulieu, Christian
Department département
Biomedical Engineering
Organization id
9
Institution établissement
University of Alberta
Province en
Alberta
Province fr
Alberta
Country en
CANADA
Country fr
CANADA
Fiscal year exercice financier
2015
Competition year année de concours
2014
Award amount
42000
Program id
RGPIN
Program name en
Discovery Grants Program - Individual
Program name fr
Programme de subventions à la découverte - individuelles
Group en
Discovery grants programs
Group fr
Programmes de subventions à la découverte
Committee code
1510
Committee name en
Electrical and Computer Engineering
Committee name fr
Génie électrique et informatique
Area of application code
518
Area of application group en
Manufacturing processes and products
Area of application group fr
Procédés et produits industriels
Area of application en
Medical equipment and apparatus
Area of application fr
Appareils et instruments médicaux
Research subject code
1900
Research subject group en
Biomedical engineering
Research subject group fr
Génie biomédical
Research subject en
Biomedical engineering
Research subject fr
Génie biomédical
Partie
2
Nb partie
5
Application title
Technical Development of Multinuclear Sodium Magnetic Resonance Imaging
Keyword
NSERC
Application summary
Engineering advances over the last 30 years have made magnetic resonance imaging (MRI) a powerful imaging modality for the non-invasive investigation of the human body. Routine clinical MRI measures signal from the hydrogen (1H) nuclei in water. However, with the appropriate hardware and imaging techniques, MRI can also acquire signal from other nuclei such as sodium (23Na) and potassium (39K), ions which may be more specific to aspects of tissue injury and disease than the ubiquitous water molecule. The ability to image sodium can provide a measure of cartilage degradation in osteoarthritis, a debilitating condition affecting millions of Canadians. Potassium is a key ion in brain function, but methods to measure it in human brain are lacking. The purpose of this grant is to develop highly sensitive radiofrequency (RF) coil hardware (e.g. sensitive phased arrays) and imaging methodology to enable quantitative imaging of these nuclei either in cartilage (23Na) or brain (39K). This proposal builds on our decade's worth of sodium MRI research, which produced world-leading images of human brain and then knee (funded over last 5 years by NSERC).
Sodium (and potassium) MRI is very challenging because of low concentration in tissue, small magnetogyric ratio, rapid signal decay, complex spin physics, and the need for nucleus-specific hardware and optimal methods. Our previous sodium MRI research on cartilage of the knee focused on the design of optimal acquisition strategies, but we used only standard volume RF coils. However, more complex and sensitive phased-array RF coils, which consist of many small localized elements, can dramatically increase signal-to-noise ratio and thus yield major gains in image resolution and quantification. While phased-array technology is now standard for regular 1H MRI, it is in its infancy for sodium MRI and needs to be explored not only for knee, but also for other body regions such as wrist, ankle, hip, and spine (where cartilage is also affected by osteoarthritis). Potassium has even greater technical challenges to overcome than sodium, and there are only two recent preliminary reports of potassium MRI in human brain.
The novel MRI hardware and methods will be designed, simulated, constructed, programmed, tested, and published by the trainees. The developments will be on a `triple strength' high field 4.7T MRI to enable major gains in signal. We hypothesize that our technical MRI advances will enable the accurate and precise measurement of sodium in cartilage and potassium in brain.
Specific Aims:
1) To develop phased-array receive-only sodium RF coils and dual-frequency (23Na/1H) detunable, concentric transmit volume coils for imaging the knee, ankle, and wrist. In this case the volume coil provides uniform excitation and the phased-array provides high receive sensitivity. The capability for simultaneous reception from both the phased-array and volume coils using counter-rotating-current coil elements will be explored to facilitate accelerated image intensity correction related to the spatially varying RF sensitivity profiles of the phased-array elements.
2) To develop transceive dual-frequency phased-array RF coils for sodium MRI of the spine and hip, where uniform volume excitation is not feasible for sodium. An associated aim is to develop and optimize multiple coil element transmission for uniform excitation.
3) To develop quantitative potassium MRI of human brain.
After 5 years, we will have developed novel multi-nuclear radiofrequency coil hardware and optimized MRI acquisition to enable imaging of (i) sodium in cartilage of the extremities, hip, and spine which are key structures affected in osteoarthritis, and (ii) potassium in the brain, a novel biomarker of brain metabolism.